High pressure on 2695 when pumping - WKB74790
Article number: 74790
SYMPTOMS
- Fluidics pressure is higher than usual.
- The pressure is high even if the column is removed.
- The pressure is not high with wet prime and purge injectors.
ENVIRONMENT
- 2695 Separations Module
CAUSE
High pressure on the tubing at the injector OUT side (in front of column)
FIX or WORKAROUND
- Replace the seal pack assembly
ADDITIONAL INFORMATION
・ C18SM7911A or later, D18SM4962A or later: Performance PLUS Seal Pack Replacement Kit (Part Number 700011784)
・ C18SM7911A or earlier, D18SM4962A or earlier: Seal Pack Assembly (Part Number 700002791)
Replace the inline filter (WKB3561)