Article number: 3561

SYMPTOMS

  • Pressure higher than usual without a column
  • Wet prime pressure is >1,000psi
  • Plunger homing overpressure errors on initialization

ENVIRONMENT

  • 2695 Separations Module
  • Alliance Separations Module

CAUSE

Partially blocked inline filter.

FIX or WORKAROUND

  1. Replace the inline filter insert assy WAT088084.

ADDITIONAL INFORMATION

 