Leaks visible on the drip tray, and a "leak detected" message for ACQUITY UPLC component - WKB71301
Article number: 71301
SYMPTOMS
- Visible leaks on the drip tray under the leak sensor
- "Leak detected" message in the console
ENVIRONMENT
- ACQUITY UPLC
- ACCQUITY UPLC H-Class
- ACQUITY UPLC I-Class
- ACQUITY ARC
- ACQUITY UPLC M-Class
- ACQUITY Premier
CAUSE
Detected leak.
FIX or WORKAROUND
- Remove and dry the sensor.
- Dry the sensor reservoir.
- Reinstall the sensor.
- Reset the detector from the ACQUITY Console.
- Check for any loose fittings.
- Repeat the run.
ADDITIONAL INFORMATION
ACQUITY UPLC Evaporative Light Scattering Detector Getting Started Guide - See section "Maintaining the leak sensor"
id71301, UPELS, UPELSARC