ACQUITY H-Class SM-FTN shows low UV area counts and shifting RTs - WKB71345

Article number: 71345

SYMPTOMS

  • Lower than expected peak area counts
  • Shifting RTs
  • QSM leak test passes
  • System leak test, excluding the needle and needle seal, passes
  • Characterize needle seal passes
  • Needle seal readiness test passes
  • Sample syringe leak test passes
  • System leak test, including the needle and needle seal, fails

ENVIRONMENT

  • ACQUITY H-Class SM-FTN

CAUSE

Faulty needle and Vespel seat

FIX or WORKAROUND

Replace sample needle and vespel seat (Parts Locator)

Replacing just vespel seat (Parts Locatormay resolve the issue