Skip to main content
Waters

ACQUITY H-Class SM-FTN shows low UV area counts and shifting RTs - WKB71345

Article number: 71345

SYMPTOMS

  • Lower than expected peak area counts
  • Shifting RTs
  • QSM leak test passes
  • System leak test, excluding the needle and needle seal, passes
  • Characterize needle seal passes
  • Needle seal readiness test passes
  • Sample syringe leak test passes
  • System leak test, including the needle and needle seal, fails

ENVIRONMENT

  • ACQUITY H-Class SM-FTN

CAUSE

Faulty needle and Vespel seat

FIX or WORKAROUND

Replace sample needle and vespel seat (Parts Locator)

Replacing just vespel seat (Parts Locator) may resolve the issue

ADDITIONAL INFORMATION


Not able to find a solution? Click here to request help.