ACQUITY H-Class SM-FTN shows low UV area counts and shifting RTs - WKB71345
Article number: 71345
SYMPTOMS
- Lower than expected peak area counts
- Shifting RTs
- QSM leak test passes
- System leak test, excluding the needle and needle seal, passes
- Characterize needle seal passes
- Needle seal readiness test passes
- Sample syringe leak test passes
- System leak test, including the needle and needle seal, fails
ENVIRONMENT
- ACQUITY H-Class SM-FTN
CAUSE
Faulty needle and Vespel seat
FIX or WORKAROUND
Replace sample needle and vespel seat (Parts Locator)
Replacing just vespel seat (Parts Locator) may resolve the issue
ADDITIONAL INFORMATION