ACQUITY UPLC BSM shows leak in drip tray, loss of eluent without pumping - WKB71279
Article number: 71279
SYMPTOMS
- Leak in drip tray
- Loss of eluent without pumping
- Sometimes "leak detected" error
ENVIRONMENT
- ACQUITY UPLC BSM
CAUSE
Fitting of eluent inlet tube on degasser is not tight enough.
FIX or WORKAROUND
- Remove the eluent inlet tube.
- Replace the fittings as needed, or cut the end of the tube.
- Tighten the tube on the degasser inlet.
ADDITIONAL INFORMATION
id71279, mobile phase, UPBINARY