Error: Sample pressure low; possibly no wash solvent or needle damaged - WKB71204
Article number: 71204
SYMPTOMS
- When attempting to perform an injection with an ACQUITY fixed-loop sample manager, the error message "Sample pressure low; possibly no wash solvent or needle damaged" appears in the ACQUITY console
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC I-Class Sample Manager Fixed-Loop (SM-FL)
- ACQUITY UPC2 Sample Manager-Fixed Loop
CAUSE
Bent or damaged sample needle or damaged wash syringe.
FIX or WORKAROUND
- Replace the sample needle
- Check wash syringes for any damage or obvious leaks.
- Calibrate the X,Y and Zp axes (UPLC SM-FL) or Calibrate R axis (UPLC I-Class SM-FL and µSM-FL)
- Calibrate the needle z axis
- Characterize the needle seal and the needle and sample loop volumes
- Ensure that the sample manager can successfully perform injections
ADDITIONAL INFORMATION
Follow this link for video showing how to remove the needle assembly and to install a replacement.