CO2 inlet pressure is too low on ACQUITY UPC2 - WKB71120

Article number: 71120

SYMPTOMS

  • CO2 pressure varies
  • CO2 inlet pressure is 0-1 psi
  • ABPR Health test fails

ENVIRONMENT

  • ACQUITY UPC2
  • UPC2 Convergence Manager

CAUSE

The CO2 tank valve is not open.

The CO2 tank is empty.

FIX or WORKAROUND

  1. Verify that the CO2 tank valve is open.
  2. Replace the CO2 tank.

ADDITIONAL INFORMATION