Can Nitrogen Cylinder be used on ELSD (2424) - WKB67898

Article number: 67898

ENVIRONMENT

  • 2424 Evaporative Light Scattering Detector

ANSWER

We don't recommend the use of cylinders with the 2424 detector. Only Nitrogen generators are recommended and supported configuration 

ADDITIONAL INFORMATION

Note: On a Standard J size cylinders, the customer may only get 40 hours of usage at normal flow rates.
The Nitrogen tank/generator must be about 2 Meter or closer to the detector other wise there is a risk of flooding the detector due to luck of enough supply of Nitrogen from the generator/tank 