ACQUITY UPLC Sample Organizer "Transfer shuttle move error", 3rd solution - WKB67739

Article number: 67739

SYMPTOMS

  • Error: Transfer shuttle move error

ENVIRONMENT

  • ACQUITY UPLC Sample Organizer

CAUSE

Sample plate is jammed in the organizer.

FIX or WORKAROUND

  1. Power-off the sample organizer.
  2. Remove the jammed plate.
  3. Power-on the sample organizer.
  4. Run Scan and Store from the sample organizer ACQUITY console tab.

ADDITIONAL INFORMATION

 

id67739, UPBINARY