ACQUITY PDA fails SQT drift test - WKB67378

Article number: 67378

SYMPTOMS

  • Drift above specification for SQT

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC Tunable UV Detector
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC
  • ACQUITY UPLC I-Class

CAUSE

Contaminated water

FIX or WORKAROUND

  1. Replace mobile phase A with fresh Milli-Q water.
  2. Prime solvent lines for 5 minutes.
  3. Rerun test injections and the noise and drift test.

ADDITIONAL INFORMATION

Water left on a system for as little as 24 hours may adversely impact the noise and drift test.