SM-FTN "Rotary Tray Home H/W Fault" - WKB67391

Article number: 67391

SYMPTOMS

  • "Rotary Tray Home H/W Fault"
  • Needle tubing bent

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

Needle tubing snagged on vials. Most likely the sample plate was not inserted into the SM-FTN properly.

FIX or WORKAROUND

Replace the SM-FTN needle with a needle appropriate for the configuration.

ADDITIONAL INFORMATION