"String" error message in console for Sample Manager - FTN - WKB67352
Article number: 67352
SYMPTOMS
- Error message in console: "String"
ENVIRONMENT
- ACQUITY UPLC H-Class FTN
- ACQUITY UPLC H-Class PLUS FTN
CAUSE
Communication problem with console
FIX or WORKAROUND
Replace the 3V battery (700001834). The Snaphat battery 700001834 is obsolete and no longer available. That CPU board can be replaced with a later version using a different battery. To replace the board contact your local service organization.
ADDITIONAL INFORMATION
id67352, A-30SM, comms, UPISMFTN, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO