Skip to main content
Waters

"String" error message in console for Sample Manager - FTN - WKB67352

Article number: 67352

SYMPTOMS

  • Error message in console: "String"

ENVIRONMENT

  • ACQUITY UPLC H-Class FTN
  • ACQUITY UPLC H-Class PLUS FTN

CAUSE

Communication problem with console

FIX or WORKAROUND

Replace the 3V battery (700001834). The Snaphat battery 700001834 is obsolete and no longer available. That CPU board can be replaced with a later version using a different battery. To replace the board contact your local service organization.

ADDITIONAL INFORMATION

 

id67352, A-30SM, comms, UPISMFTN, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.