With H-Class FLR, poor resolution of peaks for glycan/oligonucleotide methods - WKB67344

Article number: 67344

SYMPTOMS

  • Chromatogram shows merged peaks as opposed to individual peaks

ENVIRONMENT

  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC FLR Detector

CAUSE

The active preheater is set to active, disabled.

FIX or WORKAROUND

Set the active preheater to active.

ADDITIONAL INFORMATION

Console > Sample manager/column manager > Configure > Preheaters > [choose "active"]

Also ensure that the instrument method states either "console" or "active".

id67344, A-30FL, APH, UPFLRARC, UPFLRDET, UPPFLR, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO