When normalizing a 2475 with water, what emission wavelength is expected for the Raman peak? - WKB66735
Article number: 66735
ENVIRONMENT
- Waters 2475 FLR
ANSWER
397 nm
ADDITIONAL INFORMATION
From the 2475 Operators Manual:
If the Raman peak for water is within 3 nm of 397 nm,
the normalized units are embedded in the detector’s memory. If not, the detector does not
save the values but instead preserves the prior constants.