ACQUITY Arc QSM-R exhibits no flow and no pressure - WKB65495

Article number: 65495

SYMPTOMS

  • ACQUITY Arc QSM-R immediately encounters system overpressure
  • Multiple cycles of solvent-line priming produce no solvent at the waste outlet.

ENVIRONMENT

  • ACQUITY Arc Quaternary Solvent Manager (QSM-R)
  • ACQUITY Quaternary Solvent Manager

CAUSE

Check valves at either the primary or the accumulator pump are clogged or ceased.

Loose, deformed or missing check valve PEEK washer.

clipboard_e251fc88585fb0aa1974c55384537adee.png

FIX or WORKAROUND

Replace the check valve or valves as necessary. Replace check valve PEEK washer (700005221).

ADDITIONAL INFORMATION

700010664 - ACQUITY Arc QSM-R Primary Check Valve (1/pk)

700005164 - ACQUITY, ASSY, Check Valve, Double Ball and Seat (1/pk) (Arc QSM & H-Class QSM)

700005165 - ACQUITY i2 Valve cartridge, H-Class

700005221 - ACQUITY, Washer, Check Valve, PEEK