Skip to main content
Waters

Agilent LC instrument failure when setting up method using a system with a G1330 or G7130- WKB65503

Article number: 65503

SYMPTOMS

  • Newly created Agilent LC system in Empower
  • "Thermostat Installed" and "Thermostat Controlled by Method" are selected in the Pre-Configuration Utility for the Autosampler
  • Newly created instrument method

ENVIRONMENT

  • Empower 3
  • Agilent Instrument Control Framework (ICF) Support Version 3.0

CAUSE

Temperature mode is not set in the control panel on the Run Samples screen, as instructed in TECN134909208.

FIX or WORKAROUND

Set the Temperature Mode to Variable in the control panel for the Autosampler.

ADDITIONAL INFORMATION

Using the Agilent G1330A/B Autosampler Thermostat with Empower - TECN134909208

id65503, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.