Solvent leaks from injector on Alliance 2695 Separations Module - WKB64373

Article number: 64373

SYMPTOMS

  • Solvent leaks from base of seal pack assembly
  • Solvent could be mobile phase or needle wash

ENVIRONMENT

Alliance HPLC System

2695 Separations Module

CAUSE

Leak is caused by the seal pack assembly.

FIX or WORKAROUND

  1. Rebuild the seal pack.

2.   Perform the adjust seal diagnostic.

3.   Perform a compression check.

ADDITIONAL INFORMATION

New seal pack assembly part numbers:


700011784 - Kit, Performance Plus Seal pack replacement

700002791 - Spare, Seal pack assembly, 2695