nanoACQUITY system over pressure - WKB6436
Article number: 6436
SYMPTOMS
- High pressure error with the flow going
ENVIRONMENT
- nanoACQUITY UPLC
- ACQUITY UPLC M-Class
CAUSE
The nano Tee is blocked.
FIX or WORKAROUND
Sonicate the nano-T in methanol/water to clear the blockage, or replace the Tee.
ADDITIONAL INFORMATION
The part number for the nano Tee, M-M-M detail is 289002576 for the nanoACQUITY and 700009920 for the ACQUITY M-Class.