Regeneration pump 1525µ stops flow during the run - WKB64350 Last updated Save as PDF Share Share Tweet Share Article number: 64350 SYMPTOMS Regeneration pump 1525µ pump switches off the flow during the run Randomly, every few samples, the flow profile drops to baseline Changing the run time may change the time interval at which the flow stops occur Degasser LED may be flashing ENVIRONMENT 1525µ Binary HPLC Pump CAUSE A faulty degasser pump times out without displaying any error in the software or the pump. FIX or WORKAROUND Replace the degasser pump. ADDITIONAL INFORMATION