Split chromatographic peaks on ACQUITY UPLC systems - WKB64383

Article number: 64383

SYMPTOMS

  • Chromatographic peaks are split or are missing

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY I-Class SM-FL
  • ACQUITY I-Class SM-FTN
  • ACQUITY H-Class SM-FTN

CAUSE

Injection valve screw is loose and over time the valve has moved slightly forward.

FIX or WORKAROUND

  1. Tighten the injection valve screw.
  2. If needed, replace the injection valve, which might be damaged.

ADDITIONAL INFORMATION