Loss of prime on Alliance 2695 when running a gradient method - WKB6212
Article number: 6212
SYMPTOMS
- Loss of prime
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
- 2795 Separations Module
CAUSE
Running above 2 mL/minute
FIX or WORKAROUND
- Replace the GPV tubing with the High-flow GPV inlet tubing (WAT270968) when running at 2 mL/min or higher.
ADDITIONAL INFORMATION
WAT270986 2690 GPV - PRIMARY HD TUBE (.030 SS) Standard Flow SS tubing in black Teflon tubing
WAT270976 2690 LO FLOW IN TUBE (.020 SS)
WAT270968 2690 HIGH FLOW INLET TUBING (.040 SS)
id6212, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, gradient proportioning valve