Over pressure error on ACQUITY UPLC Sample Manager - WKB60062

Article number: 60062

SYMPTOMS

  • Over pressure error on ACQUITY UPLC Sample Manager with column removed
  • POD correctly fitted

ENVIRONMENT

  • ACQUITY UPLC Sample Manager 

CAUSE

Blocked column stabilizer

FIX or WORKAROUND

Replace the column stabilizer (205000489).

ADDITIONAL INFORMATION