What switch position is used to identify with and control the column selection valve in the Alliance 30cm Column Heater Cooler? - WKB5997
Article number: 5997
ENVIRONMENT
- Alliance Series Column Heater Module Alliance Series Column Heater/Cooler Module.
ANSWER
Alliance 30cm Column Heater and 30cm Column Heater/Cooler - Column Selection Valve Options.
- Three different column selection valves are available as an option. You must configure valves using the switch located on the back of the column module:
- Position 0 - No valve
- Position 1 - Three Column Select Valve (205000162)
- Position 7 - Two Column Regeneration valve (205000163)
- Position F - Six Column Select Valve (205000164)
ADDITIONAL INFORMATION
id5997, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB