What switch position is used to identify with and control the column selection valve in the Alliance 30cm Column Heater Cooler? - WKB5997

Article number: 5997

ENVIRONMENT

  • Alliance Series Column Heater Module
  • Alliance Series Column Heater/Cooler Module

ANSWER

Alliance 30cm Column Heater and 30cm Column Heater/Cooler - Column Selection Valve Options

  • Three different column selection valves are available as an option
  • You must configure valves using the switch located on the back of the column module:
    • Position 0 - No valve
    • Position 1 - Three Column Select Valve (205000162)
    • Position 7 - Two Column Regeneration valve (205000163)
    • Position F - Six Column Select Valve (205000164)

 

ADDITIONAL INFORMATION

 