Skip to main content
Waters

Bent needle on 2707 SM - WKB59518

Article number: 59518

SYMPTOMS

  • Bent needle

ENVIRONMENT

  • 2707 Sample Manager

CAUSE

Needle alignment caused the needle to hit the vial cap

FIX or WORKAROUND

  • Replace the needle (70003842 - Standard Needle Assembly, 15µL).

Directions to replace the needle are in Section 3.6 of the 2707 Autosampler Overview and Maintenance Guide (71500168202).

  • Calibrate needle positions

ADDITIONAL INFORMATION

Other options:

Preparative Needle Assembly, 60 µL (700004085)
Bio-Compatible Needle Assembly, 15 µL (700003843)

Not able to find a solution? Click here to request help.