Baseline noises with 2414 detector - WKB59481
Article number: 59481
SYMPTOMS
- Baseline noise with 2414 Refractive Index Detector
- There are no pressure ripples of pump flow
- Baseline noise didn't occur in stop flow
ENVIRONMENT
- 2414 Refractive Index Detector
CAUSE
3-way Recycle Valve (part number 700002362) block.
FIX or WORKAROUND
- Remove the tube at the entrance of the 3-way Recycle Valve (lower valve) that connects from the 2-Way Solenoid Valve (valve upper left)
- If baseline noise is gone, blockage in 3-way Recycle Valve
- Replace the 3-way Recycle Valve