Blocked injection pod on 2777 Sample Manager with DLW upgrade - WKB58815

Article number: 58815

SYMPTOMS

  • Variability in ISTD peak areas
  • Black deposits observed at injection pod

ENVIRONMENT

  • 2777 Sample Manager with DLW upgrade

CAUSE

Blocked injection valve

FIX or WORKAROUND

Replace the blocked injection valve (700011083).

ADDITIONAL INFORMATION