Air bubbles in the sample syringe of an ACQUITY UPLC Sample Manager and UPLC H-Class FTN - WKB5880
Article number: 5880
SYMPTOMS
- Air bubbles in the sample syringe
- Lost of sensitivity
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC H-Class FTN
CAUSE
Unknown
FIX or WORKAROUND
- Ensure that there is adequate weak wash/purge solvent in the reservoir.
- Prime the sample syringe from the ACQUITY Console. (Move on to step 3 if priming doesn't resolve the issue.)
- Replace the sample syringe and re-prime (if bubbles still remain move to step 4.)
- Disconnect the line coming from the valve manifold feeding the sample syringe.
- Place the line in a vial of weak wash and prime syringe, if the sample syring primes up you have a faulty valve manifold, replace with part #700010171.
ADDITIONAL INFORMATION
For sample syringe part numbers, see article What are the part numbers for the ACQUITY Sample Manager Flow Through Needle sample syringes? - WKB10580