Air bubbles in the sample syringe of an ACQUITY UPLC Sample Manager and UPLC H-Class FTN - WKB5880

Article number: 5880

SYMPTOMS

  • Air bubbles in the sample syringe
  • Lost of sensitivity

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC H-Class FTN

CAUSE

Unknown

FIX or WORKAROUND

  1. Ensure that there is adequate weak wash/purge solvent in the reservoir.
  2. Prime the sample syringe from the ACQUITY Console. (Move on to step 3 if priming doesn't resolve the issue.)
  3. Replace the sample syringe and re-prime (if bubbles still remain move to step 4.)
  4. Disconnect the line coming from the valve manifold feeding the sample syringe.
    1.  Place the line in a vial of weak wash and prime syringe, if the sample syring primes up you have a faulty valve manifold, replace with part #700010171.

ADDITIONAL INFORMATION

For sample syringe part numbers, see article What are the part numbers for the ACQUITY Sample Manager Flow Through Needle sample syringes? - WKB10580

id5880, UPBINARY