Error: Autozero offset excessive on 2414 Refractive Index detector - WKB5831
Article number: 5831
SYMPTOMS
- Autozero offset excessive observed during system wash
- Baseline not stable
ENVIRONMENT
- 2414 Refractive Index (RI) detector
CAUSE
Contents of the flow cell.
FIX or WORKAROUND
- Purge detector with fresh solvent
- Take purge mode off and flush detector
- Reboot
ADDITIONAL INFORMATION
When using a new column, do not flow to detector until proper conditioning of column is complete. Otherwise, extended multi-day purging may be necessary.