ACQUITY UPLC Column manager is leaking, but no leak is visible - WKB57972
Article number: 57972
SYMPTOMS
- Leak sensor triggered for the column manager
- No visible leak from any of the four columns
- Column pressure is stable
- Leaking occurs regardless of the column position
- Overall system pressure is less than normal
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC Column Manager
CAUSE
The column manager outlet-to-detector inlet tubing (port 7) has come loose or wear and tear.
FIX or WORKAROUND
Replace the column manager outlet-to-detector inlet tubing (see Additional Information for part numbers).
ADDITIONAL INFORMATION
700004163 - Column Manager Outlet Tubing, 0.004" ID (Black PEEK)
700003442 - Column Manager Outlet Tubing, 0.0025" ID (Tan PEEK)