nanoACQUITY UPLC fails to calibrate system volume - WKB57974

Article number: 57974

SYMPTOMS

  • nanoACQUITY UPLC fails to calibrate system volume
  • System volumes are correct
  • Wash solvent composition is correct

ENVIRONMENT

  • nanoACQUITY UPLC

CAUSE

Leak on the sample loop

FIX or WORKAROUND

  1. Refit the sample loop.
  2. Prime wash solvents.
  3. Calibrate system volume.

ADDITIONAL INFORMATION