The gold fitting broke and is stuck in the vent valve of ACQUITY UPLC I-Class BSM - WKB57342

Article number: 57342

SYMPTOMS

  • Gold fitting broke and is stuck in the vent valve

ENVIRONMENT

  • ACQUITY UPLC I-Class BSM
  • ACQUITY UPLC Classic BSM

CAUSE

Over-tightened fitting

FIX or WORKAROUND

  1. Attempt to remove the fitting without damaging the vent valve.
  2. If unsuccessful, replace the vent valve cartridge.

ADDITIONAL INFORMATION