Alliance system pressure fluctuates in water/aqueous conditions - WKB57411
SYMPTOMS
- 
    Alliance pressure fluctuations with high delta (~200 psi at 1500 psi) in aqueous conditions with any composition of water - 
        Instrument has had PM and rebuild several times 
- 
        Pump tray swapped and GPV replaced without resolution 
- 
        Flow rate accuracy unstable 
 
- 
        
- 
    100% line C in methanol at 1 mL/min, the system is operating and stable at ~500+ psi with delta of ~10 psi - 
        100% line D in methanol gives same stable operating conditions 
- 
        Flow rate accurate in methanol 
 
- 
        
- 
    Swapping lines between water and methanol, the observed pressure fluctuations move with aqueous mobile phase - 
        Water mobile phase is freshly prepared (18 meg ohm purity) 
- 
        Degasser operating correctly at 0.5 psia - 
            Degassed by vacuum filtration and sonicated 
 
- 
            
 
- 
        
ENVIRONMENT
- Alliance 2695 Separations Module
CAUSE
Microbial growth and contamination of the system's wetted surfaces
FIX or WORKAROUND
- Passivate the system with 30% phosphoric acid.
ADDITIONAL INFORMATION
id57411, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent, gradient proportioning valve, pressure ripple
 
 