Degasser reported H/W fault on ACQUITY Solvent Manager - WKB5732

Article number: 5732

SYMPTOMS

  • Error message "Degasser reported H/W Fault"

ENVIRONMENT

  • nanoACQUITY UPLC
  • UPLC
  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-ClassQSM
  • ACQUITY UPLC  I-Class BSM
  • ACQUITY QSM-R

CAUSE

False error message

FIX or WORKAROUND

-Power cycle LC and restart acquisition computer.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.