2487 error: Peak Not Found - Reference Energy Maximum - WKB55343
Article number: 55343
SYMPTOMS
- "Peak Not Found. Reference Energy Maximum" error appears at startup
- Intensities or peak response of known peaks are less than expected
ENVIRONMENT
- 2487 Dual-Wavelength Absorbance Detector
- 2489 UV/Vis Detector
CAUSE
Lamp intensity is too low.
FIX or WORKAROUND
- Replace the lamp (WAS081142) if 2487, if you have 2489 use 201000281.
- Refer to manual for the replacement procedure.