Wash syringe fails leak test on ACQUITY UPLC Sample Manager- Fixed Loop- WKB55349
Article number: 55349
SYMPTOMS
- Wash syringe fails the leak test
- Both or one of the wash syringes fails the Wash syringe leak test
- All other tests pass
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Fixed Loop
- ACQUITY UPLC I-Class
- ACQUITY UPC2 System
CAUSE
Leaking wash syringe
FIX or WORKAROUND
- Replace the wash syringe.
- Thoroughly prime the wash syringes prior to run the Wash syringe leak test.
ADDITIONAL INFORMATION
See article How to replace a wash syringe on an ACQUITY Sample Manager
ACQUITY UPLC Sample Manager Wash Syringe, 2.5 mL 700002569
ACQUITY UPLC I-CLass/UPC2 Fixed Loop Wash Syringe, 2.5ml, Inverted 700006070