Wash syringe fails leak test on ACQUITY UPLC Sample Manager- Fixed Loop- WKB55349

Article number: 55349

SYMPTOMS

  • Wash syringe fails the leak test
  • Both or one of the wash syringes fails the Wash syringe leak test
  • All other tests pass 

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Fixed Loop
  • ACQUITY UPLC I-Class
  • ACQUITY UPC2 System

CAUSE

Leaking wash syringe

FIX or WORKAROUND

  1. Replace the wash syringe.
  2. Thoroughly prime the wash syringes prior to run the Wash syringe leak test.

ADDITIONAL INFORMATION

See article How to replace a wash syringe on an ACQUITY Sample Manager

ACQUITY UPLC Sample Manager Wash Syringe, 2.5 mL 700002569

ACQUITY UPLC I-CLass/UPC2 Fixed Loop Wash Syringe, 2.5ml, Inverted  700006070