No data on Alliance HPLC System - WKB53093
Article number: 53093
SYMPTOMS
- No data in Run Samples window
- No data (data missing) in Review
- Injection starts and run clock starts normally
- Eventual time-out with Instrument Failure
ENVIRONMENT
- Empower 3 Enterprise
- Alliance HPLC System
- 2695 Separations Module
- 2998 Photodiode Array Detector
CAUSE
Missing Inject Start trigger cable to detector
FIX or WORKAROUND
Fit a trigger cable (part WAT020312) from the Alliance (InjectStart) to the detector (InjectStart).