No data on Alliance HPLC System - WKB53093
Article number: 53093
SYMPTOMS
- No data in Run Samples window
- No data (data missing) in Review
- Injection starts and run clock starts normally
- Eventual time-out with Instrument Failure
ENVIRONMENT
- Empower 3 Enterprise
- Alliance HPLC System
- 2695 Separations Module
- 2998 Photodiode Array Detector
CAUSE
Missing Inject Start trigger cable to detector
FIX or WORKAROUND
Fit a trigger cable (part WAT020312) from the Alliance (InjectStart) to the detector (InjectStart).
ADDITIONAL INFORMATION
id53093, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, 2998, 2998B, 2998SFC, A-10PDA, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB