No data on Alliance HPLC System - WKB53093

Article number: 53093

SYMPTOMS

  • No data in Run Samples window
  • No data (data missing) in Review
  • Injection starts and run clock starts normally
  • Eventual time-out with Instrument Failure

ENVIRONMENT

  • Empower 3 Enterprise
  • Alliance HPLC System
  • 2695 Separations Module
  • 2998 Photodiode Array Detector

CAUSE

Missing Inject Start trigger cable to detector 

FIX or WORKAROUND

Fit a trigger cable (part WAT020312) from the Alliance (InjectStart) to the detector (InjectStart).

ADDITIONAL INFORMATION

 