Poor RSD on Alliance HPLC System with 2695 Separations Module - WKB53082
Article number: 53082
SYMPTOMS
- Poor RSD
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Worn seal pack
FIX or WORKAROUND
- Prime and flush the system.
- Perform the seal pack adjustment diagnostic.
- If issue persists, rebuild the seal pack.
- Seal Pack Rebuild Kit w/Needle - WAT271019
- Seal Pack Rebuild Kit w/Needle - 700011783 (for new installed enhancement kit or serial no after 2018)
ADDITIONAL INFORMATION
For seal pack adjustment refer to page 61 of the Waters Alliance e2695 Separations Module Operator's Guide - 715003794.
https://www.waters.com/waters/suppor...1556&type=USRM
For seal pack rebuild refer to page 141 through 146 of the Operator's Guide.
If rebuilding the seal pack doesn't work replace the seal pack assembly and the restrictor loop (part number: 700002791).