"Organizer configuration modified" error on ACQUITY UPLC Sample Organizer - WKB53073

Article number: 53073

SYMPTOMS

  • "Organizer configuration modified" error seen for ACQUITY UPLC Sample Organizer

  • Rebooting PC and modules does not resolve issue

ENVIRONMENT

  • MassLynx
  • ACQUITY UPLC

CAUSE

The sample organizer detected a change in configuration.

FIX or WORKAROUND

  1. Remove all sample plates from the sample organizer.

  2. In the ACQUITY console, go to Sample Organizer > Configure > Scan and Store Shelf Layout, and perform the function twice.

  3. Return the sample plates to the sample organizer and click the "Verify" button in the Sample Organizer console. 

ADDITIONAL INFORMATION

id53073, UPBINARY, UPCPORG, UPORGANIZE, UPORGIVD