"Organizer configuration modified" error on ACQUITY UPLC Sample Organizer - WKB53073
Article number: 53073
SYMPTOMS
-
"Organizer configuration modified" error seen for ACQUITY UPLC Sample Organizer
-
Rebooting PC and modules does not resolve issue
ENVIRONMENT
- MassLynx
- ACQUITY UPLC
CAUSE
The sample organizer detected a change in configuration.
FIX or WORKAROUND
-
Remove all sample plates from the sample organizer.
-
In the ACQUITY console, go to Sample Organizer > Configure > Scan and Store Shelf Layout, and perform the function twice.
-
Return the sample plates to the sample organizer and click the "Verify" button in the Sample Organizer console.