Seal wash leak on 2695 - WKB53010
Article number: 53010
SYMPTOMS
- Seal wash leaks in the seal wash pump head connection
- Air bubbles seen in the seal wash line to the left of a pump head, but not on the right side
ENVIRONMENT
- 2695 Separations Module
- 2795 Separations Module
CAUSE
The seal wash tube has bad connection to the pump head.
FIX or WORKAROUND
Reseat the seal wash tubes on the pump head.

ADDITIONAL INFORMATION
Trim the ends of the tubing if they are compressed/ damaged