SYMPTOMS

  • Seal wash leaks in the seal wash pump head connection

ENVIRONMENT

  • 2695 Separations Module
  • 2795 Separations Module

CAUSE

The seal wash tube has bad connection to the pump head.

FIX or WORKAROUND

Reseat the seal wash tubes on the pump head.

clipboard_eadbe9460eacee9d498c839eb35ba4a72.png

ADDITIONAL INFORMATION

Trim the ends of the tubing if they are compressed/ damaged