High C/D ratio observed on 2695 - WKB53036
SYMPTOMS
- High C/D ratio observed on 2695
- Erratic baseline
- Pressure fluctuations
ENVIRONMENT
- 2695
- Empower
CAUSE
The solvent reservoir filter is blocked, causing pump cavitation during flow.
Not enough back pressure is being applied.
Leaking primary check valve and/or high-pressure plunger seal and/or face seal
Leaking fitting between the primary outlet and the inlet to the check valve cartridge
FIX or WORKAROUND
Replace the solvent reservoir filter (WAT025531).
ADDITIONAL INFORMATION
Note: A column or restriction needs to be used to help remove dissolved air; help the action of the check valves; and help the plunger seals to flex and seal correctly on the plunger. Perform a static leak test.