High C/D ratio observed on 2695 - WKB53036
Article number: 53036
SYMPTOMS
- High C/D ratio observed on 2695
- Erratic baseline
- Pressure fluctuations
ENVIRONMENT
- 2695 Separations Module
- 2695e Separations Module
CAUSE
The solvent reservoir filter is blocked, causing pump cavitation during flow.
ADDITIONAL INFORMATION
Note: A column or restriction needs to be used to help remove dissolved air; help the action of the check valves; and help the plunger seals to flex and seal correctly on the plunger.