Alliance seal-wash pump makes no sound - WKB5290
Article number: 5290
SYMPTOMS
- Seal-wash pump is making no sound
- Unable to pump seal wash
ENVIRONMENT
- Alliance 2690
- Alliance 2695
- Alliance 2790
- Alliance 2795
CAUSE
Defective seal-wash solenoid
FIX or WORKAROUND
- Replace seal-wash pump (WAT270926).
- If seal wash pump is still not making noise, it could be a faulty PCB. Replace pump driver PCB (WAT270919).
- Verfiy functionality by performing seal wash priming.
ADDITIONAL INFORMATION
A Waters trained engineer may need to replace defective parts.
For the seal-wash pump solenoid, see the Graphical Parts Locator link WAT270926.
Contact your local Waters office.