When changing the needle on an ACQUITY I-Class with Fixed Loop Sample Manager, which axis must be calibrated? - WKB4891
Article number: 4891
ENVIRONMENT
- ACQUITY I-Class
- Sample Manager - Fixed Loop
ANSWER
When changing the needle, calibrate the needle's z axis, and then characterize the needle seal. This procedure can be found in the ACQUITY UPLC Sample Manager - Fixed Loop Operator's Overview and Maintenance Information.
ADDITIONAL INFORMATION
