When changing the needle on an ACQUITY I-Class with Fixed Loop Sample Manager, which axis must be calibrated? - WKB4891

Article number: 4891

ENVIRONMENT

  • ACQUITY I-Class
  • Sample Manager - Fixed Loop

ANSWER

When changing the needle, calibrate the needle's z axis, and then characterize the needle seal. This procedure can be found in the ACQUITY UPLC Sample Manager - Fixed Loop Operator's Overview and Maintenance Information.

ADDITIONAL INFORMATION

 