Infinity Series 2 stack does not show up in DHCP - WKB4676
Article number: 4676
SYMPTOMS
- Infinity Series 2 stack does not show up in DHCP
ENVIRONMENT
- Infinity Series 2 on Empower network
CAUSE
The dip switch on the rear of G7117 A,B,C DAD needs setting.
FIX or WORKAROUND
- As with all Ethernet Agilent systems, Waters recommends hooking up the Ethernet cable to the detector. In this case, the dip switch needs setting.
- On the rear of the detector, notice the six-position bank of dip switches.
- Viewing from left to right, dip switch 4 must be up, all others down.
- Reboot the PC and detector. The system will now show in DHCP.
- Give the system a unique name and create a new system.