ENVIRONMENT

  • ACQUITY UPLC PDA Detector

ANSWER

If the detector has a serial number starting with A18 or higher, do NOT downgrade the firmware. These detectors contain a new PDA device that requires its native firmware. If the firmware is downgraded, there will be no signal whatsoever, resulting in a filter wheel "failure to home" error.

ADDITIONAL INFORMATION

 

id46769, A-30PDA, UPAPDA, UPBINARY, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA