How to remove the collet from a compressed reusable high-pressure fingertight fitting - WKB3756

Article number: 3756

OBJECTIVE or GOAL

Remove the collet from a compressed reusable high-pressure fingertight fitting.

ENVIRONMENT

  • ACQUITY column stabilizer

PROCEDURE

1. Order Part Number 700003170 for the "MULTI-TOOL, COLLET & COMP. SCREW, HPFT".

2. Use the tool to remove the collet from the compression screw.

Disassemble collet.JPG

3. Replace the PEEK ferrule (700003114 for the FERRULE, PEEK, 1/16, HPFT, PKG 10).

ADDITIONAL INFORMATION

Part Number compression screw 700003169 (10 pack)

Part Number for collet 700003168 (10 pack)