"Leak detected" error with ACQUITY UPLC Column Heater - WKB3416
Article number: 3416
SYMPTOMS
- Leak detected error in the ACQUITY console
ENVIRONMENT
- ACQUITY UPLC Column Heater
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
Leak is detected in the column compartment.
FIX or WORKAROUND
- Connection from active pre-heater to column is not properly connected.
- Inspect for fluid at the leak sensor; clean if necessary.
- Reset the sample manager.
- If leak persists, replace the column and active pre-heater.
ADDITIONAL INFORMATION
id3416, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO