Error: Sample Syringe HW Lost Steps (value) on ACQUITY Sample Manager - WKB3198

Article number: 3198

SYMPTOMS

  • Sample Syringe HW lost Steps (56) error in Console on ACQUITY UPLC
  • Unable to prime sample syringe/perform injections

ENVIRONMENT

  • ACQUITY UPLC System
  • ACQUITY Sample Manager (SM)

CAUSE

  • Worn/damaged metering syringe drive timing belt.
  • Damaged metering syringe drive

FIX or WORKAROUND

  1. Power-off Sample Manager
  2. Try manually turning the belt at the base of the sample syringe.
  3. Replace worn/loose timing belt.

ADDITIONAL INFORMATION

Timing belt part number: 700003909

Included in  ASSY, SYRINGE DRIVE, METERING - 700002581

id3198, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR