Article number: 319503

OBJECTIVE or GOAL

How to adjust the sample manager fixed loop (FL) O ring

ENVIRONMENT

  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS

PROCEDURE

  • Open the sample compartment door and fluidics compartment door.
  • Using the T20 TORX driver, loosen the two captive screws that secure the access panel.

clipboard_5d203bd0-ace1-42ae-b733-36d496d43203.png

  •  Insert a piece of used PEEK needle or PEEK tubing into the needle wash seal port, and gently push the needle down through the O-ring. You should feel slight resistance as the needle slides through the O-ring.

clipboard_15e78784-a712-428d-bfd7-2cc7f7cc0a26.png

  •  If you do not feel slight resistance while sliding the needle through the O-ring, the needle wash seal port is probably too tight or too loose. Manually adjust the needle wash seal port until you can verify that it is seated correctly. (CAUTION: TO AVOID DAMAGING THE O-RING SEAL AND NEEDLE, DO NOT OVER-TIGHTEN THE NEEDLE WASH SEAL PORT).
  •  Remove the PEEK needle or PEEK tubing from the needle wash seal port.

ADDITIONAL INFORMATION


 