Skip to main content
Waters

How to adjust the O-ring in an I-Class or ACQUITY UPC2 SM-FL - WKB319503

Article number: 319503

OBJECTIVE or GOAL

Adjust the Sample Manager - Fixed Loop (SM-FL) O-ring.

ENVIRONMENT

  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPC2 System
  • ACQUITY Classic

PROCEDURE

  • Open the sample compartment door and the fluidics compartment door.
  • Using a T20 TORX driver, loosen the two captive screws that secure the access panel.

clipboard_5d203bd0-ace1-42ae-b733-36d496d43203.png

  • Insert a piece of used PEEK needle or PEEK tubing into the needle wash seal port, and gently push the needle down through the O-ring. You should feel slight resistance as the needle slides through the O-ring.

clipboard_15e78784-a712-428d-bfd7-2cc7f7cc0a26.png

  • If you do not feel slight resistance while sliding the needle through the O-ring, the needle wash seal port is probably too tight or too loose. Manually adjust the needle wash seal port until you can verify that it is seated correctly. (CAUTION: TO AVOID DAMAGING THE O-RING SEAL AND NEEDLE, DO NOT OVER-TIGHTEN THE NEEDLE WASH SEAL PORT).
  •  Remove the PEEK needle or PEEK tubing from the needle wash seal port.

ADDITIONAL INFORMATION


 

Not able to find a solution? Click here to request help.