No baseline with 2475 Fluorescence detector - WKB3192
Article number: 3192
SYMPTOMS
- Low lamp hours
- Failed to Normalize, gives error peak location not near 397 nm. Is there water in cell? Values not saved (40100).
ENVIRONMENT
- ACQUITY FLR
- 2475 detector (Fluorescence detector)
CAUSE
Detector saturated.
FIX or WORKAROUND
- Run fresh HPLC-grade methanol through flowcell for 15 minutes.
- Run fresh HPLC water through flowcell for 15 minutes.
- Reboot the detector.
- Perform manual wavelength calibration (Shift+3, also see manual).
- Perform normalization per 2475 manual. If you have an ACQUITY FLR, see the following screen shot:
Link to 2475 operator's guide / manual
Link to ACQUITY FLR guide
ADDITIONAL INFORMATION
id3192, 2475, 2475B, A-10FL