Agilent 6890 GC with G1888 Headspace doesn't make second injection - WKB2996

Article number: 2996

SYMPTOMS

  • Instrument previously worked correctly

ENVIRONMENT

  • Empower
  • Agilent 6890 GC
  • G1888 Headspace

CAUSE

 

FIX or WORKAROUND

  1. On the G1888 Headspace front panel, go to Menu > Advanced > Enable RS232, and select No.

  2. Reboot the GC, the headspace, and then the LAC/E.

ADDITIONAL INFORMATION

 